ZCLED-42A02\N
Mix color tester 20V/2000mA
Basic function
Basic function
- Electric characteristics: VF, DVF, VFD, VZ and IR
- Optical characteristics: LOP, λp, λd, λc, hw, purity, (x,y), CCT, CRI
- To avoid misjudgment, it is kelvin measurement and has the contact resistance function.
- Polarity determination and preheating function.
- It'sworkable for linking to automatic machine.
- It can connect with ESD. (Optional.)
- It's available for 4-dices in single package and it also has mix color test.可
- The device must have one pin in common.
Specific function
- The minimum of VF is 0.1uA.
- The maximum of VR is 200V.
- HBM mode is 8000V when linking to ESD.
- Data analysis. (LEDDispensing software only.)
- Yield observation and searching data. (LEDPreRealTime software only.)
- I-V curve function. (LED software only.)
- Operation is easy.
SPCIFICATION
Test item |
Reading |
Output item |
Test condition |
Remark |
||||||
Unit |
Range |
Accuracy | Reproducibility |
Unit | Range |
Accuracy | Output time |
|||
VF | V | 0.1~20.00 | ±0.25%+0.002V | ±0.02 | IF | mA | 0.001~4.000 | ±1%+0.002 | 2~99mS | 1.2.8 |
4.01~40.00 | ±1%+0.02 | 0.3~99mS | ||||||||
40.1~400.0 | ±1%+0.2 | 0.3~99mS | ||||||||
401~2000 | ±1%+2 | 0.3~99mS | ||||||||
DVF | V | 0.1~20.00 | ±0.25%+0.002V | ±0.02 | DIF | mA | 0.001~4.000 | ±1%+0.002 | 2~99mS | 1.2.8 |
4.01~40.00 | ±1%+0.02 | 0.3~99mS | ||||||||
40.1~400.0 | ±1%+0.2 | 0.3~99mS | ||||||||
401~2000 | ±1%+2 | 0.3~99mS | ||||||||
POLA | Find polarity, no test reading |
POLA IF |
mA | 4.01~400.0 | ±1%+0.2 | 1mS | ||||
HEAT | Source only, no test reading |
HEAT IF |
mA | 0.001~4.000 | ±1%+0.002 | 2~99mS | 1.2.8 | |||
4.01~40.00 | ±1%+0.02 | 0.3~99mS | ||||||||
40.1~400.0 | ±1%+0.2 | 0.3~99mS | ||||||||
401~2000 | ±1%+2 | 0.3~99mS | ||||||||
VFD | V | 0.1~20.00 | Proximity |
±0.1 | IFD | uA | 100uA | Proximity |
1.2.8 | |
mA | 25mA | |||||||||
VZ | V | 0.1~200.0 | ±0.25%+0.2 | ±0.2 | IZ | uA | 10.00~40.00 | ±1%+0.02 | 2~99mS | 1.2.8 |
40.1~400.0 | ±1%+0.2 | 2~99mS | ||||||||
IR | uA | 0.001~4.000 | ±1%+0.002uA | ±0.005uA | VR | V | 0.01~200.0V | ±0.5%+0.02V | 2~99mS | 1.2.8 |
10.00~40.00 | ±1%+0.02uA | ±0.05uA | ||||||||
40.1~400.0 | ±1%+0.2uA | ±0.5uA | ||||||||
LOP (Iv or Ie) |
0.01~4.000 | ±2% | ±1% | LOP IF |
mA | 0.001~4.000 | ±1%+0.002 | 2~99mS | 1.3.8 | |
4.01~40.00 | ±2% | ±1% | 4.01~40.00 | ±1%+0.02 | 0.3~99mS | |||||
40.1~400.0 | ±2% | ±1% | 40.1~400.0 | ±1%+0.2 | 0.3~99mS | |||||
Use calibration to correspond with Iv or Ie |
401~2000 | ±1%+2 | 0.3~99mS | |||||||
λd | nm | 380.0~700.0 | ±0.5nm | ±0.3nm | spec IF |
mA | 0.001~4.000 | ±1%+0.002 | 2~99mS | 4.5.6.7.8 |
λc | nm | 380.0~720.0 | ±0.5nm | ±0.3nm | 4.01~40.00 | ±1%+0.02 | 0.2~99mS | |||
λp | nm | 380.0~1000.0 | ±0.5nm | ±0.3nm | 40.1~400.0 | ±1%+0.2 | 0.2~99mS | |||
HW | nm | 3.0~600.0nm | ±0.5nm | ±0.3nm | 401~2000 | ±1%+2 | 0.2~99mS | |||
Purity | 0.000~1.000 | ±0.002 | ±0.001 | 1.CD is 2048 pixel. 2.The least ST is 0.2 mS. |
||||||
CIE x.y | 0.0002~1.0000 | ±0.001 | ±0.0005 | |||||||
CCT | °K | 1000~25000 | ±100 | ±50 | ||||||
mix color test | Test LOP and wavelength only. |
spec IFX4 |
mA | 1.0~20.00 | ±1%+0.02 | 1~99mS | 4.5.6.7.8 | |||
20.1~200.0 | ±1%+0.2 | 1~99mS | ||||||||
201~2000 | ±1%+2 | 1~99mS |
- 1.In order to get normal test result for LED, please make sure it is not affected by heating and environment.
- 2.When S/N is worse, it needs to use [50/60HZ]. The time of 50Hz needs higher than 20mS. The time of 60Hz needs higher than 16.6mS.
- 3.Calibrated with a Standard LED with spectrum mismatch adjustment.
- 4.λd, Purity and CCT are calculated from (x, y) refer to (x,y) for accuracy.
- 5.A LED with flat spectrum width has worse λp accuracy.
- 6.Using stanadrd UV lighting source to calibrate UV range of spectrum.
- 7.λd is only referable for visible light.
- 8.When current is lower, it needs to increase test time.
Dimension |
Output curve |
SPCIFICATION
Dimension | 48.5(W)*52(L)*22(H)cm |
Weight | 18.98KG |
AC input power | 110V/220V |
Watt | 732W |